Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMIL

US FDA — Medical Devices · 2025-06-26 · SwitzerlandClass I
Issuing authority
US FDA — Medical Devices
Alert date
2025-06-26
Country of alert
Switzerland
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
07630002802963
Brand on record
Hamilton Medical AG

What the authority reported

Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.