Official recall record
Breathing circuit set, coaxial, 1.80 m, with flow sensor and expiratory valve set, REF: 260128, used with HAMILTON-C1, HAMILTON-T1 and HAMIL
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2025-06-26
- Country of alert
- Switzerland
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 07630002802963
- Brand on record
- Hamilton Medical AG
What the authority reported
Ventilator coaxial breathing circuit may have a crack in blue tubing, which is used to deliver fresh gas to the patient. The crack will compromise inspiratory and expiratory flows, leading to partial or complete rebreathing of exhaled gases, which could lead to acute hypercapnia, respiratory acidosis, and if unrecognized organ dysfunction.