Official recall record
CARESCAPE Canvas 1000, Model Numbers: a) 5865770; b) 5865770-01040169; c) 5865770-01054142; d) 5865770-01054696; e) 5865770-01054698;
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-04-05
- Country of alert
- Finland
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00195278522320
- Brand on record
- GE Healthcare Finland Oy
What the authority reported
GE HealthCare has become aware of the potential for loss of monitoring on CARESCAPE Canvas 1000, CARESCAPE Canvas Smart Display, and certain versions of CARESCAPE B850 and CARESCAPE B650 monitors if, following the replacement of the CPU timekeeper battery, a mains power loss occurs during active monitoring. In the unlikely event that this occurs, it can result in the displays or monitors not turning on, after the mains power is restored. This can result in a delay in recognizing changes in patient condition until a replacement monitor is connected.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-09-26 · US FDA — Medical Devices
Portrait VSM Model nos. 6660000-666 Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms - 2025-09-26 · US FDA — Medical Devices
B105P, B125P, B105M, B125M, B155M Patient Monitors; Model nos. 6160000-002 6160000-003 6160000-004 6160000-005 - 2025-09-26 · US FDA — Medical Devices
BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manuf