Official recall record
Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, S
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-06-18
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00763000002039
- Brand on record
- Medtronic, Inc.
What the authority reported
In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 3 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-06-17 · US FDA — Medical Devices
Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software compo - 2025-06-18 · US FDA — Medical Devices
Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Cr - 2025-06-17 · US FDA — Medical Devices
Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following softwar
Other recalls from this brand
- 2026-03-11 · US FDA — Medical Devices
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment o - 2025-06-18 · US FDA — Medical Devices
Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. - 2025-06-17 · US FDA — Medical Devices
Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following soft
All 6 official recall notices naming Medtronic, Inc. on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.