ProductGuru Recalls

Official recall record

Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Sof

US FDA — Medical Devices · 2025-06-18 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-06-18
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00763000002039
Brand on record
Medtronic, Inc.

What the authority reported

In prior SmartSync application versions, the Abort button stopped the test that was selected. During an induction test, there was a limited window of time for the user to abort a therapy, thereby limiting the user's ability to cancel a high voltage therapy delivery.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 3 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 6 official recall notices naming Medtronic, Inc. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.