Official recall record
Baxter Novum IQ Large Volume Pump, Product Code REF 40700BAXUS
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-07-14
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 05413765851797
- Brand on record
- Baxter Healthcare Corporation
What the authority reported
Baxter is issuing an Urgent Medical Device Correction for the Novum IQ large volume pump (LVP) due to the potential for underinfusion when transitioning from a flow rate to a higher flow rate that is more than double. (e.g., rate change or bolus). The level of underinfusion is variable based on the current infusion rate, the duration the pump has been running at this flow rate, and the magnitude of the rate change. The longer the duration the pump has been running at the current infusion rate and the larger the magnitude of the rate change, the larger the underinfusion that would be experienced. Additionally, Baxter has identified an increase in customer reports of over and underinfusion potentially due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Consistent with the instructions for use, customers should ensure that: 1) The door is fully open before loading the set. 2) The tubing is taut and loaded without slack in the pumping channel.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-08-04 · US FDA — Medical Devices
Novum IQ LVP (Large Volume Pump), Product Code 40700BAXUS, infusion pump
Other recalls from this brand
- 2026-08-18 · US FDA — Medical Devices
PRO+ 40" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake - 2026-08-18 · US FDA — Medical Devices
PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system - 2026-08-18 · US FDA — Medical Devices
PRO+ 36" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake
All 43 official recall notices naming Baxter Healthcare Corporation on this site →