Medtronic SynchroMed Flex Infusion Mode - A810 Clinician Programmer Software Application, loaded on the following devices: 1. PROG CT900

US FDA — Medical Devices · 2026-04-08 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-04-08
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00643169890923
Brand on record
Medtronic Neuromodulation

What the authority reported

Software issue: When programming the SynchroMed Pump with the A810 CP App in Flex Infusion Mode, the infusion steps should automatically be arranged in consecutive order, from earliest to latest. However, in rare cases, the steps may be displayed out of order. If this occurs, the programmed Flex Infusion schedule will not be delivered at the intended time, if at all.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.