Contour next GEN Blood Glucose Monitoring System

US FDA — Medical Devices · 2023-05-19 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-05-19
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00301937917011
Brand on record
Ascensia Diabetes Care US, Inc.

What the authority reported

Ascensia Diabetes Care has determined through customer complaints that 580 of 2900 meters from lot number DM01T033P are reconfigured from mg/dL to mmol/L. Among them, 579 meters distributed in the United States were packaged into the meter kit and had incorrect factory-set unit of measure where the meters display glucose results in mmol/L rather than mg/dL. The standard unit of measurement for the United States is mg/dL. If a consumer does not notice the incorrect unit of measurement, it is possible that the meters glucose measurement will be read as a lower blood glucose measurement than expected, and this may result in the patients glucose level remaining high, which can potentially lead to an injury requiring immediate medical intervention.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.