ProductGuru Recalls

Official recall record

LINEAR 7.5Fr.IAB, Intra-Aortic Balloon Catheter and Accessories, used to provide counter pulsation therapy in the aorta Product Descripti

US FDA — Medical Devices · 2023-06-23 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-06-23
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Datascope Corp.

What the authority reported

During IAB catheter insertion, the introducer dilator may fracture at the hub, leaving the body of the introducer dilator housed within the sheath. Risks to health include damage to the femoral artery, descending aorta, or embolization; surgical intervention and delay of IAB therapy may also result. Firm has received 10 complaints, including 3 serious adverse events and 1 patient death.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

All 14 official recall notices naming Datascope Corp. on this site →