RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2025-06-26
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Siemens Healthcare Diagnostics Inc
What the authority reported
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-07-07 · US FDA — Medical Devices
epoc BGEM BUN Test Card [25pk]. Material Number: 10736515. - 2024-09-24 · US FDA — Medical Devices
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515 - 2024-05-17 · US FDA — Medical Devices
Software version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPID