ProductGuru Recalls

Official recall record

ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures includi (ref Z-2242-2024)

US FDA — Medical Devices · 2024-06-26 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-06-26
Country of alert
Germany
Alert level
Class II
Alert type
recall
Brand on record
Siemens AG/Siemens Healthcare GmbH

What the authority reported

A potential issue with ARTIS icono systems equipped with a small detector and a corresponding collimator (model # 10843101) was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose rate may be higher than intended by the user.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Germany · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.