Official recall record
NeuMoDx Cartridge-used for extraction, purification, amplification and detection of nucleic acids on the NeuMoDx 288 and NeuMoDx 96 Molecula
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-05-24
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 10814278020274
- Brand on record
- NeuMoDx Molecular Inc
- Product category
- Medical Device
What the authority reported
Top label on some of the NeuMoDx cartridges within these lots may have been incorrectly applied during the manufacturing process, resulting in encroachment into the PCR region. The misplaced label may interfere with the module s optics performance, thereby increasing the risk for false positive results
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2021-05-04 · US FDA — Medical Devices
NeuMoDx Cartridge, For In Vitro Diagnostic Use on the NeuMoDx 288 and NeuMoDx 96 Molecular Systems, Model Number 100100