ProductGuru Recalls

Official recall record

Neurostimulation Kit (Ported, Dual 8, 40 cm), REF: 71005-US Peripheral Nerve Stimulation Kit (Ported, Dual 8, 40 cm), REF: 71019-US Comp

US FDA — Medical Devices · 2024-06-03 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-06-03
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00812537036137
Brand on record
Nalu Medical, Inc.

What the authority reported

Neurostimulation Systems have a manufacturing non-conformance that may reduce the efficiency of the radiofrequency signals between the Implantable Pulse Generator (IPG) and the Therapy Disc, causing the IPG to be non-functional, which causes the device to be non-functional so it will no longer inhibit the transmission of pain signals to the brain.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.