Official recall record
Brand Name: Exacta Mix Product Name: Vented Micro-Volume Inlet Model/Catalog Number: H938175 Product Description: Tubing sets with pus
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-04-22
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00085412475806
- Brand on record
- Baxter Healthcare Corporation
What the authority reported
Affected lots may contain incorrect 0.8-micron filters. Use of a 0.8-micron filter in place of the validated 3-micron filter may lead to reduced flow efficiency, incomplete or slower transfers of ingredients, or system alarms. This issue may lead to clinical effects such as electrolyte imbalance, metabolic instability, delayed recovery, etc.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-09-16.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2024-08-20 · US FDA — Medical Devices
Baxter EXACTAMIX Inlet, Vented, Micro-Volume Inlet, REF H938175, for use in a pharmaceutical compounding device
Other recalls from this brand
- 2026-08-18 · US FDA — Medical Devices
PRO+ 36" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake - 2026-08-18 · US FDA — Medical Devices
PRO+ MRS SURFACE FOR VERSACARE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system - 2026-08-18 · US FDA — Medical Devices
PRO+ 40" MRS SURFACE, mattress with Advanced Microclimate Technology (AMT) uses a non-powered system of intake
All 43 official recall notices naming Baxter Healthcare Corporation on this site →