ProductGuru Recalls

Official recall record

PhaseOne Antimicrobial Solution, Model/Catalog Number: 15050

US FDA — Medical Devices · 2026-04-27 · MexicoClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-04-27
Country of alert
Mexico
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00860009268616
Brand on record
Oculus Technologies of Mexico, S.A. de C.V.

What the authority reported

The recall is being conducted due to corrosion of the internal metallic spring within the sprayer component. This condition is caused by material incompatibility between the product solution and the stainless steel spring and may result in visible discoloration and the presence of trace levels of metallic elements (including iron, chromium, and nickel) in the solution during use.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · Mexico · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.