ProductGuru Recalls

Official recall record

Philips Azurion 7M20, Model Numbers: 722079, 722224 with Software release R2.2.0, R2.2.1, R2.2.3, R2.2.5 and R2.2.6 with the Roadmap Pro So

US FDA — Medical Devices · 2023-12-07 · NetherlandsClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-12-07
Country of alert
Netherlands
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00884838099258
Brand on record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

What the authority reported

During the device phase of Roadmap Pro, a user may experience a dark image that is not clinically usable. The dark image may result after the user performs a pedal tap (shortly touching the footswitch pedal without generating an Xray image). The pedal tap may cause a dark image with the next fluoroscopy run.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · Netherlands · all filters

Same barcode in other registries

The same EAN appears in 3 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 90 official recall notices naming PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on this site →