Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad Avanta Injection System; Catalog number: AV
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-04-24
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Bayer Medical Care, Inc.
What the authority reported
Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.