Tracheal Intubation Fiberscope OLYMPUS LF-P- To be used with an Olympus Light Source, documentation equipment, display monitor and other anc
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-05-26
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 04953170051098
- Brand on record
- Olympus Corporation of the Americas
What the authority reported
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible