Official recall record
Olympus Endoscope Reprocessor OER-Mini-For use in cleaning and high-level disinfection of heat sensitive Olympus flexible endoscopes and the
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-05-26
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 04953170331619
- Brand on record
- Olympus Corporation of the Americas
What the authority reported
Revised OER-Pro, OER-Elite, OER-Mini endoscope reprocessor labeling- the LF-V and LF-P are no longer compatible endoscopes for reprocessing in the OERs. The addendum highlights the following changes: 1. Removal of LF-V and LF-P as compatible with the OER 2. Change conditions for ETO gas sterilization from 12% EtO to 100% EtO 3. Removing the reusable cleaning brush for the manual cleaning. Single Use brushes remain compatible
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.