Official recall record
Giraffe Warmer with Integrated Resuscitation System (iRes), Panda Warmer with Integrated Resuscitation System (iRes) Product Identifiers: M1
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-05-08
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00813841020010
- Brand on record
- Datex-Ohmeda Inc
What the authority reported
Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.
Record last checked against the registry on 2026-09-16.
Same barcode in other registries
The same EAN appears in 2 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2026-05-08 · US FDA — Medical Devices
Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifiers: 20826 - 2026-05-08 · US FDA — Medical Devices
Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service blender M109
Other recalls from this brand
- 2026-05-08 · US FDA — Medical Devices
Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product identifi - 2026-05-08 · US FDA — Medical Devices
Panda Freestanding Warmer, Infant Radiant Warmer Product Identifiers: 2063568-001 Impacted only if service bl
All 9 official recall notices naming Datex-Ohmeda, Inc. on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.