ProductGuru Recalls

Official recall record

UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift system. Instrument for insertion of a cystoscope shaft into the

US FDA — Medical Devices · 2023-06-23 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-06-23
Country of alert
Germany
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
04250480106617
Brand on record
Scholly Fiberoptic Gmbh

What the authority reported

There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Germany · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.