Official recall record
BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327; CII Safe ES, REF: 1116-00; CII
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-05-02
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 10885403512667
- Brand on record
- CareFusion 303, Inc.
- Product category
- Medical Device
What the authority reported
If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-11-20 · US FDA — Medical Devices
BD Pyxis MedStation ES, REF: 323, BD Pyxis MedStation ES 7 Drawer Auxiliary, REF: 324, BD Pyxis MedStation ES Tower - 2024-02-01 · US FDA — Medical Devices
BD Pyxis: MedStation ES, REF: 323; Anesthesia Station ES, REF: 327; MedStation ES Tower, REF: 352; - 2025-01-08 · US FDA — Medical Devices
(1) Pyxis MedStation ES, Catalog No. 323; (2) Pyxis Anesthesia Station (PAS), Catalog No. 327; (3) Pyxis MedStation ES - 2024-11-14 · US FDA — Medical Devices
BD Pyxis MedStation ES, REF: 323 - 2023-04-05 · US FDA — Medical Devices
BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?