MAHURKAR 12 Fr Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-06-28
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 20884521057026
- Brand on record
- Covidien LP
- Product category
- Medical Device
What the authority reported
Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus