Bicarby Dialysate; Model number: RFP-400-G;

US FDA — Medical Devices · 2026-05-07 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-05-07
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00840861102839
Brand on record
Fresenius Medical Care Holdings, Inc.

What the authority reported

There is an error in the labeling of the GTIN on the product. When scanned; however, the case labeling, lot number, and printed labeling on the product correctly identifies the product code and the solution contained in the bag. In the event the incorrect GTIN labeling is used as the singular source for product identification, unrecognized use of a potassium solution with a concentration differing from the prescription may result.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand