Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-05-23
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00380650003997
- Brand on record
- Alcon Research LLC
What the authority reported
After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-04-28 · US FDA — Medical Devices
Brand Name: Alcon Laboratories, Inc. Product Name: Constellation Vision System Vitrectomy Procedure Pak Mode - 2024-08-08 · US FDA — Medical Devices
DAILIES TOTAL1, DAILIES TOTAL1 for Astigmatism, DAILIES TOTAL1 Multifocal, TOTAL30 Sphere,
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.