ProductGuru Recalls

Official recall record

Surgical Image Guidance Functionality, software used with NGENUITY 3D Visualization System, Part Number: 8065000399

US FDA — Medical Devices · 2024-05-23 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-05-23
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00380650003997
Brand on record
Alcon Research LLC

What the authority reported

After surgery initiation, while using surgical image guidance, if significant eye rotation/movement occurs, a software anomaly may cause re-establishment of incorrect new registration angle, range indicator will display incorrectly, which could result in incorrect placement of toric IOL axis, which could cause astigmatic error under/over correction resulting in decreased uncorrected visual acuity.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.