Official recall record
ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-05-31
- Country of alert
- Denmark
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 05700693938004
- Brand on record
- Radiometer Medical ApS
What the authority reported
Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2020-05-26 · US FDA — Medical Devices
Blood gases (PCO2, PO2) and blood pH test system; ABL800 FLEX; Model Number 393-800; UDI (01)05700693938004. Model 393-8
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.