ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood

US FDA — Medical Devices · 2024-05-31 · DenmarkClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-05-31
Country of alert
Denmark
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
05700693938004
Brand on record
Radiometer Medical ApS

What the authority reported

Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.