Official recall record
Brand Name: Perfusor Space Product Name: Perfusor Space Syringe Pump, Wireless Model/Catalog Number: 8713031U Software Version: 8713031U
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-06-14
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 4046964797958
- Brand on record
- B Braun Medical Inc
- Product category
- Medical Device
What the authority reported
Perfusor Space Syringe Pump's instructions for use previously identified Monoject syringes as an option for use. However, changes to manufacturing to Cardinal Health Monoject syringes may result in issues with recognition, compatibility, and pump performance, such as overdose, underdose, delay in therapy, delay in occlusion alarms, and delay in feeding. Cardinal Health Monoject syringes are not compatible with the indicated pumps.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-05-06 · US FDA — Medical Devices
Spinal Needle procedure kits: Material Description (Material Number) 24GA. PENCAN SPINAL BUP. TRAY W/DURAPR - 2026-05-06 · US FDA — Medical Devices
Spinocan Spinal Needle procedure kits: Material Description (Material Number): Spinocan 26 Ga. x 3-1/2 in. - 2026-05-06 · US FDA — Medical Devices
Design Options Spinal Tray Spinal Needle Anesthesia procedure kits: Material Description (Material Number):
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?