ProductGuru Recalls

Official recall record

Micro Joystick R-net. Electrical wheelchair component.

US FDA — Medical Devices · 2025-07-10 · BelgiumClass I
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-07-10
Country of alert
Belgium
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
05407008320027
Brand on record
mo-Vis BVBA

What the authority reported

Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Belgium · all filters

Other recalls from this brand