Official recall record
Micro Joystick R-net. Electrical wheelchair component.
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-07-10
- Country of alert
- Belgium
- Alert level
- Class I
- Alert type
- recall
- Barcode (EAN/GTIN)
- 05407008320027
- Brand on record
- mo-Vis BVBA
What the authority reported
Devices with firmware versions 2.3 and lower may experience timing variations between software tasks, which could result in skipping the necessary neutral check and the wheelchair producing unintended movement upon power-up when the joystick is out of neutral.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2025-07-10 · US FDA — Medical Devices
IDM-MICRO-R. Electrical wheelchair component.
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.