ProductGuru Recalls

Official recall record

Ritmed Neurological Sponges, 0.75x0.75in, Sterile, REF: 85995C

US FDA — Medical Devices · 2026-05-08 · CanadaClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-05-08
Country of alert
Canada
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10686864044465
Brand on record
AMD Medicom Inc.
Product category
Medical Device

What the authority reported

Neurological sponges used to moisten or absorb biological fluid during neurological surgery and may also be used to protect the brain during incision, may not consistently meet endotoxin specification for devices that contact the cerebrospinal fluid. Elevated endotoxin levels may lead to neurovascular adverse events, febrile response, and/or local transient inflammation, hypotension, or nausea.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · Canada · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.