ProductGuru Recalls

Official recall record

GS70H Salus Surgical Table, Product Code 18-071-70

US FDA — Medical Devices · 2023-08-01 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-08-01
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Skytron, LLC

What the authority reported

When the control of the operating table was used in Bluetooth mode, the operating table continued moving even after releasing the button on the control. Originally, the root cause was thought to be confined to Bluetooth/wireless communications between the control pendant and the surgical table. However, during investigation, Mizuho discovered that the failure could be replicated even in wired mode if communication between the pendant and table were somehow interrupted. The root cause appears to be: if communication between the controller and table is interrupted while a button is pressed, the table does not know if/when a button is released. This occurs in wireless or wired mode.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.