LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Bat

US FDA — Medical Devices · 2023-07-07 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-07-07
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00190376228037
Brand on record
Alphatec Spine, Inc.

What the authority reported

Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.