LIF AMP, Adjustable Awl, REF 117-165, Part of the AMP System. Used with Spine Lateral Interbody Systems, IdentiTi LIF, Transcend LIF and Bat
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-07-07
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00190376228037
- Brand on record
- Alphatec Spine, Inc.
What the authority reported
Awl instrument adjustable drill button assembly assembled in the incorrect orientation, preventing locking feature from engaging with shaft at the desired set point; instruments lack ability to control awl depth by means of adjustable shaft advancement stop, which may lead to over insertion, dural tear, vascular/neurologic injury, adjacent tissue damage, increased operative time, revision surgery.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-05-13 · US FDA — Medical Devices
SafeOp Patient Interface, REF: AIX11240, part of SafeOp 3: Neural Informatix System
ProductGuru Recalls republishes official safety information with attribution. We are not