SIGNA Victor, with affected software versions: MR30.1; Nuclear Magnetic Resonance Imaging System
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-06-14
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- GE Medical Systems, LLC
What the authority reported
GE HealthCare has become aware that for certain MR systems, the system predicted B1+RMS value can exceed the B1+RMS user prescribed limit when scanning in Low SAR Mode and the following two conditions are met: 1. 2D FSE T2 FLAIR OR T2 FLAIR Propeller imaging sequence is selected, AND 2. the Optimized T2 FLAIR sequence option is turned off OR not included in the MR configuration. If this occurs, it can result in overheating of an MR conditional implant.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-06-05 · US FDA — Medical Devices
Optima XR200amx, Mobile Digital-Ready Radiographic System - 2026-06-05 · US FDA — Medical Devices
Optima XR240amx, Mobile Digital Radiographic System - 2026-06-05 · US FDA — Medical Devices
Optima XR220amx, Mobile Digital-Ready Radiographic System
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.