Link OptiStem Rasp Handle-Used with the OptiStem Rasp to prepare the femoral medullary canal for a knee prosthesis implant during a total kn
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-07-26
- Country of alert
- Germany
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 04026575284658
- Brand on record
- Waldemar Link GmbH & Co. KG (Mfg Site)
What the authority reported
A crack in the weld of the rasp handle may lead to a fracture of the frontal sleeve when the rasp is further impacted or extracted. The rasp remains intact. The rasp would need to be pulled out by a clamp leading to a prolongation of surgery or modified surgery