Official recall record
Incisive CT, software version 5.0
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-08-02
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Philips Healthcare
What the authority reported
Philips has identified three software issues with compliance concerns to Subchapter J: Issue 1: Coronary Computed Tomography Angiography (CTA) scan may not be triggered at the optimal time. Issue 2: Use of aborted surview for scan planning will result in deviation from intended scan area. Issue 3: Actual clinical scan with bariatric (Noah) couch will deviate from the intended scan area if surview scan length is greater than 1832mm.