Discrete photometric chemistry analyzer for clinical use.

US FDA — Medical Devices · 2024-06-14 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-06-14
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Beckman Coulter, Inc.

What the authority reported

Beckman Coulter, Inc. has identified a rise in reagent carousel motion errors in Access 2 Immunoassay System Analyzers that appear to be caused by misaligned index wheels housed within the reagent carousel module. Customers affected by this issue may observe an increase in reagent carousel motion errors or QSMC errors that occur while unloading or loading reagent packs during system initialization and sample processing, which causes a non-operational system status and cancels tests; and patient results may be delayed if these errors cancel test processing.