GE Healthcare AW Server 3.2 ext.6.5; Radiological Image Processing System

US FDA — Medical Devices · 2026-05-08 · FranceClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-05-08
Country of alert
France
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00840682102384
Brand on record
GE Medical Systems SCS

What the authority reported

GE HealthCare has become aware of a context synchronization issue in AW Server 3.2 ext. 6.5. When a user selects a patient or exam in the AW Server Web Client worklist and launches an interactive application (e.g. Volume Viewer), the application may open the previous patient's exam instead of the intended one. When this issue occurs, there is no system warning or error notification. If the error is not recognized, a clinical user could review, interpret, or report images for the wrong patient which could lead to misdiagnosis, incorrect clinical decisions, resulting in delayed or incorrect treatment.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · France · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 6 official recall notices naming GE Medical Systems, SCS on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.