ProductGuru Recalls

Official recall record

Allura Xper FD20 Biplane OR Table Product Codes: (1)722020, (2) 722025; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K1419

US FDA — Medical Devices · 2025-09-03 · NetherlandsClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-09-03
Country of alert
Netherlands
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00884838059108
Brand on record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

What the authority reported

Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Netherlands · all filters

Same barcode in other registries

The same EAN appears in 3 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 73 official recall notices naming PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.