Official recall record
exoplan version 3.1 Rijeka Software -A medical software, intended to support the pre-operative planning of dental implants using the visuali
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2023-08-17
- Country of alert
- Germany
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 4260521365026
- Brand on record
- Exocad GmbH
- Product category
- Medical Device
What the authority reported
A software library filtering error has been discovered which occurs when users are using the Step-by-Step Full Drill Protocol export functionality of exoplan 3.1 Rijeka. When a guided surgery treatment approach is selected, along with a Step-by-Step Full Drill Protocol export implant library, exoplan software does not filter out unsupported sleeve height positions for a particular sub-full drill protocol and instead shows ALL possible sleeve height positions for all sub-full drill protocols. The use of an improper sleeve height position in a surgical guide could result in patient injury
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.