Official recall record
ZAP-X Radiosurgery System, Model: 300150
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-07-25
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00860183001504
- Brand on record
- Zap Surgical Systems
What the authority reported
If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.
What to do