therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic muta

US FDA — Medical Devices · 2023-08-22 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-08-22
Country of alert
Germany
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
04053228002048
Brand on record
Qiagen GmbH

What the authority reported

Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.