Official recall record
MAGNETOM Vida Fit. Model Number: 11410481.
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-08-28
- Country of alert
- United States
- Alert level
- Class I
- Alert type
- recall
- Brand on record
- Siemens Medical Solutions USA, Inc
What the authority reported
There is a potential for an ice blockage to form or currently exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed venting paths, leading to a pressure build-up within the helium containment system. This pressure build-up could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.