ProductGuru Recalls

Official recall record

OmniTract Surgical Retractor System, which includes 633 of the firm s product codes.

US FDA — Medical Devices · 2026-06-05 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-06-05
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
10381780268468
Brand on record
Integra LifeSciences Corp. (NeuroSciences)
Product category
Medical Device

What the authority reported

The current Omni-Tract Surgical Retractor System IFU states a dry time of 30 minutes. Integra LifeSciences has identified that the dry time needs to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.