ProductGuru Recalls

Official recall record

Optima XR240amx, Mobile Digital Radiographic System

US FDA — Medical Devices · 2026-06-05 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-06-05
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00840682122351
Brand on record
GE Medical Systems, LLC

What the authority reported

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.