ProductGuru Recalls

Official recall record

INFINION CX LEAD KIT, 50CM, Model: SC-2317-50, Material: M365SC2317500; INFINION CX LEAD KIT, 70CM, Model: SC-2317-70, Material: M365SC231

US FDA — Medical Devices · 2026-06-17 · United StatesClass I
Issuing authority
US FDA — Medical Devices
Alert date
2026-06-17
Country of alert
United States
Alert level
Class I
Alert type
recall
Barcode (EAN/GTIN)
08714729861614
Brand on record
Boston Scientific Neuromodulation Corporation

What the authority reported

Spinal cord stimulation system lead body may be exposed to stress at the anchor site. This may result in high impedance measurements or lead fractures, which may lead to inadequate stimulation, additional intervention, including lead explant and replacement.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · United States · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.