ProductGuru Recalls

Official recall record

HardyDisk AST Cefiderocol 30µg (FDC30) REF Z9435 HardyDisk AST Cefiderocol 30µg (FDC30) for in vitro agar diffusion susceptibility test

US FDA — Medical Devices · 2026-06-08 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-06-08
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Hardy Diagnostics

What the authority reported

Due to incorrect cartridges include in affected products that there is the potential for false-susceptible or false-resistant susceptibility results and may lead to selection of an ineffective antimicrobial therapy, potentially resulting in delayed appropriate treatment, prolonged infection, or worsening clinical outcome.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.

More recalls: US FDA — Medical Devices · United States · all filters

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.