Brand Name: COAPTIUM CONNECT Product Name: COAPTIUM CONNECT Model/Catalog Number: NCOCO5010 Software Version: N/A Product Description: I

US FDA — Medical Devices · 2026-06-15 · FranceClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-06-15
Country of alert
France
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
03760279380179
Brand on record
TISSIUM SA

What the authority reported

Currently labeled shelf life of 36 months is not substantiated by real-time data. Devices will be relabeled to reflect revised shelf life of 26 months. The use of a medical device beyond its validated shelf life may result in perioperative breakage of the device rendering device unusable; worst-case outcome would be revision surgery.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.