ProductGuru Recalls

Official recall record

Philips Azurion 7M20 systems with FlexArm ceiling-mounted system, Interventional Fluoroscopic X-Ray System, Model Numbers 722079 and 722224

US FDA — Medical Devices · 2024-08-06 · NetherlandsClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-08-06
Country of alert
Netherlands
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00884838085268
Brand on record
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.

What the authority reported

The motorized longitudinal movement of the FlexArm stand may be inconsistent (not smooth) and eventually may become unavailable. This issue is due to grease leakage from the bearings and/or anti-corrosion oil applied on the bearings, because of which the lubrication on the ceiling rail may be excessive, reducing the friction between the rails and the friction wheel.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Netherlands · all filters

Same barcode in other registries

The same EAN appears in 4 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.