Official recall record
TD Navio Maxi Speech Generating Device Model/Catalog Number: 1000799
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2026-07-08
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Tobii Dynavox Llc
What the authority reported
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-17.
Record last checked against the registry on 2026-09-17.
Other recalls from this brand
- 2026-07-08 · US FDA — Medical Devices
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.