Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long-Term Dual Lumen Hemodialysis Ca

US FDA — Medical Devices · 2024-07-10 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-07-10
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00801741085451
Brand on record
Bard Peripheral Vascular Inc

What the authority reported

Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.