Atellica IM CA 19-9 (250 Test - REF)-In vitro diagnostic use in the quantitative measurement of the CA 19-9 tumor-associated antigen in huma

US FDA — Medical Devices · 2024-07-26 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-07-26
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00630414293516
Brand on record
Siemens Healthcare Diagnostics, Inc.

What the authority reported

Positive bias for some samples around the CA 19-9 upper limit of normal of 35 U/m as compared to the previous lot. The sample bias observed was greatest in the Asia Pacific sample population. The observed bias is not proportional across the assay measuring interval and is less pronounced for sample results above a concentration of 110 U/mL.