Ingenia Ambition X - For use as a diagnostic device to obtain cross-sectional images, spectroscopic images and/or spectra of the internal st

US FDA — Medical Devices · 2024-09-03 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-09-03
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00884838098350
Brand on record
Philips North America Llc

What the authority reported

Integrated Radio Frequency (IRF) Carrier assembly screws may come loose and protrude beyond the surface could cause the screws to interfere with other moving parts, potentially blocking the horizontal tabletop. May lead to delay in diagnosis and/or may include anxiety of a patient is left on the tabletop inside the bore for a longer duration than anticipated. movement.